Image quality of bedside chest radiographs in intensive care beds with integrated detector tray: A phantom study
1University Hospital Southwest Denmark, Department of Radiology and Nuclear Medicine, Finsensgade 35, 6700 Esbjerg, Denmark.
Introduction:
Vendors offer intensive care beds with integrated detector trays for bedside radiography, promoting better ergonomics and patient comfort. However, no documentation of the effects on diagnostic image quality has been located. This study examines measured and subjective image quality of supine bedside chest radiographs with and without use of such a detector tray.
Methods:
A contrast-detail phantom (CDRAD 2.0) was exposed using standard supine chest exposure parameters. Plexiglass plates of 16 and 21 cm were placed in front to simulate patient attenuation for standard and adipose patients. Exposures were repeated with the detector placed in tray and directly in bed. Images were analysed using dedicated software giving a figure-of-merit IQFinv. Results were compared using ANOVA. Then an anthropomorphic chest phantom (Lungman) was exposed using the same parameters, and the same placements of the detector. Exposures were done with and without extra conformal tissue to simulate varying patient sizes, and with and without added typical intensive care equipment. Images were analysed by two radiologists using a three-point scale, on five image quality criteria. Radiologist also stated whether the images were sufficient for diagnosis. Results were compared using Visual Grading Characteristics, using dedicated software, resulting in Areas Under the Curve (AUC-VGC) for each combination and criteria. Inter- and intra-rater reliability were assessed with kappa statistics. Composite Visual Grading Analysis (VGAS) scores were calculated for each image. Both IQFinv and were normalized and compared.
Results:
For all criteria both IQFinv and AUC-VGC was significantly better when exposing the detector directly in bed, than with the detector placed in the tray across all exposures. When stratified into thin and adipose patients, IQFinv decreased significantly for thin patients, while VGAS-scores did not. For adipose patients, both figures were significantly lower with the detector in the tray.
Conclusion:
Use of detector tray for bedside chest imaging decreases image quality.
Implications For Practice:
Radiographers should critically evaluate image quality and experimentally determine optimal exposure factors, when taking equipment with integrated trays into use.
Related Concept Videos
Radiological Investigation I: X-ray and CT
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...