Simulation study on flow rate accuracy of infusion pumps in vibration conditions during emergency patient transport
Kwan Young Hong1, You Yeon Kim2, Seung Yeon Yoo1
1Department of Anesthesiology and Pain Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, 81 Irwon-ro, Gangnam-gu, Seoul, 06351, Korea.
Abstract:
Infusion pumps are frequently used when transferring critically ill patients via patient transport cart, ambulance, or helicopter. However, the performance of various infusion pumps under these circumstances has not been explored. The aim of this study was to evaluate the flow rate accuracy of infusion pumps under various clinical vibration conditions. Experiments were conducted with four different types of pumps, including two conventional syringe pumps (Injectomat MC Agilia, Fresenius Kabi and TE-331, Terumo), one conventional peristaltic pump (Volumed μVP7000; Arcomed), and one new cylinder pump (H-100, Meinntech). The flow rate was measured using an infusion pump analyzer on a stable table (0 m/s2) for 1 h with 1 ml/h and 5 ml/h. Experiments were repeated in mild vibration (2 m/s2) (representing vibration of patients in a moving stretcher or ambulance), and in moderate vibration (6 m/s2) (representing vibration in helicopter transport). Any accidental bolus occurrence in extreme vibration situations (20 m/s2) was also analyzed. Simulated vibrations were reproduced by a custom-made vibration table. In the resting state without vibration and in mild vibration conditions, all pumps maintained good performance. However, in moderate vibration, flow rates in syringe pumps increased beyond their known error ranges, while flow rates in peristaltic pumps remained stable. In extreme vibration, accidental fluid bolus occurred in syringe pumps but not in peristaltic pumps. The newly developed cylinder pump maintained stable performance and was unaffected by external vibration environments.
Related Concept Videos
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Typical Model Studies
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...