Abstract

Insights

The alpha defensin lateral flow test shows excellent performance in diagnosing periprosthetic joint infection (PJI). This diagnostic accuracy study confirms its reliability, leading to FDA authorization as a PJI diagnostic aid.

Area of Science:

  • Orthopedics
  • Infectious Diseases
  • Diagnostic Medicine

Background:

  • Periprosthetic joint infection (PJI) diagnosis relies on accurate testing.
  • Synovial fluid alpha defensin tests have shown promise for PJI detection.

Purpose of the Study:

  • To evaluate the diagnostic accuracy of a novel lateral flow test for synovial fluid alpha defensin.
  • To compare the lateral flow test's performance against a laboratory-based alpha defensin test for PJI.

Main Methods:

  • A diagnostic accuracy study was conducted using a prospective multicenter cohort and a supplemental control cohort.
  • The index test was a lateral flow immunoassay for synovial fluid alpha defensin.
  • The reference standard was the 2013 Musculoskeletal Infection Society (MSIS) multicriteria definition of PJI.

Main Results:

  • The alpha defensin lateral flow test demonstrated high sensitivity (89.5%) and specificity (94.8%) in the prospective cohort.
  • In the supplemental cohort, sensitivity was 98.5% and specificity was 98.2%.
  • No significant difference in diagnostic performance was observed compared to the alpha defensin enzyme-linked immunosorbent assay (ELISA).

Conclusions:

  • The alpha defensin lateral flow test exhibits solid diagnostic performance for PJI.
  • The test has received U.S. Food and Drug Administration (FDA) authorization.