Analytical performance of lateral flow immunoassay for SARS-CoV-2 exposure screening on venous and capillary blood
Margaret A Black1, Guomiao Shen1, Xiaojun Feng1
1Department of Pathology, NYU Langone Health, New York City, NY, United States of America.
Objectives:
We validate the use of a lateral flow immunoassay (LFI) intended for rapid screening and qualitative detection of anti-SARS-CoV-2 IgM and IgG in serum, plasma, and whole blood, and compare results with ELISA. We also seek to establish the value of LFI testing on blood obtained from a capillary blood sample.
Methods:
Samples collected by venous blood draw and finger stick were obtained from patients with SARS-CoV-2 detected by RT-qPCR and control patients. Samples were tested with Biolidics 2019-nCoV IgG/IgM Detection Kit lateral flow immunoassay, and antibody calls were compared with ELISA.
Results:
Biolidics LFI showed clinical sensitivity of 92% with venous blood at 7 days after PCR diagnosis of SARS-CoV-2. Test specificity was 92% for IgM and 100% for IgG. There was no significant difference in detecting IgM and IgG with Biolidics LFI and ELISA at D0 and D7 (p = 1.00), except for detection of IgM at D7 (p = 0.04). Capillary blood of SARS-CoV-2 patients showed 93% sensitivity for antibody detection.
Conclusions:
Clinical performance of Biolidics 2019-nCoV IgG/IgM Detection Kit is comparable to ELISA and was consistent across sample types. This provides an opportunity for decentralized rapid testing and may allow point-of-care and longitudinal self-testing for the presence of anti-SARS-CoV-2 antibodies.