An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Mechanical Evaluation of the Stability of One or Two Miniscrews under Loading on Synthetic Bone
Andrea Pradal1, Ludovica Nucci2, Nicola Derton1
1Department of Medical, Surgical and Health Sciences, University of Trieste, 34127 Trieste, Italy.
Abstract:
The aim of the present study was to evaluate the primary stability of a two-miniscrew system inserted into a synthetic bone and to compare the system with the traditional one. Forty-five bi-layered polyurethane blocks were used to simulate maxillary cancellous and cortical bone densities. Samples were randomly assigned to three groups-one-miniscrew system (Group A, N = 23), two-miniscrew system (Group B, N = 22) and archwire-only (Group C, N = 10). A total of 67 new miniscrews were subdivided into Group A (23 singles) and Group B (22 couples). 30 mm of 19″ × 25″ archwires were tied to the miniscrew. The load was applied perpendicularly to the archwire. Maximum Load Value (MLV), Yield Load (YL) and Loosening Load (LL) were recorded for each group. The YL of Group B and C had a mean value respectively of 4.189 ± 0.390 N and 3.652 ± 0.064 N. The MLV of Group A, B and C had a mean value respectively of 1.871 ± 0.318N, of 4.843 ± 0.515 N and 4.150 ± 0.086 N. The LL of Group A and B had a mean value respectively of 1.871 ± 0.318 N and of 2.294 ± 0.333 N. A two- temporary anchorage device (TAD) system is on average stiffer than a one-TAD system under orthodontic loading.
More Related Videos
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
07:09In situ Compressive Loading and Correlative Noninvasive Imaging of the Bone-periodontal Ligament-tooth Fibrous Joint
Published on: March 7, 2014