Quantitative positron emission tomography imaging in the presence of iodinated contrast media using electron density
Nadav Shapira1, Joshua Scheuermann1, Amy E Perkins2
1Department of Radiology, University of Pennsylvania, Philadelphia, PA, USA.
Purpose:
As preparation for future positron emission tomography (PET)/dual-energy computed tomography (DECT)T imaging modality and new possible clinical applications, the study aimed to evaluate the utility of clinically available spectral results from a DECT system for improving attenuation corrections of PET acquisitions in the presence of iodinated contrast media. The dependence of the accuracy of PET quantification values, reconstructed with conventional and spectral-based attenuation corrections, was examined as a function of the amount of iodine content and x-ray radiation exposure.
Methods:
Measurements were performed on commercial PET/CT and DECT systems, using a semi-anthropomorphic phantom with seven centrifuge tubes in its bore. Five different configurations of tube contents were scanned by both PET/CT and DECT. With the aim of mimicking clinically observed concentrations, in all phantom configurations the center tube contained a high concentration of radionuclide while the peripheral tubes contained a lower concentration of radionuclide. Iodine content was incrementally increased between phantom configurations by replacing iodine-free tubes with tubes that contained the original radionuclide concentration within a 10 mg/ml iodine dilution. DECT-based attenuation correction maps were generated by scaling electron density spectral results into corresponding 511 keV photon linear attenuation coefficients.
Results:
Mean SUV values obtained from the nominal PET reconstruction, using conventional CT images as input for the attenuation correction, demonstrate a monotonic increase of 8.6% when the water and radionuclide mixtures were replaced by iodine, water, and radionuclide (same level of activity) mixture. Mean SUV values obtained from the DECT-based reconstruction, in which the attenuation correction utilizes electron density values as input, demonstrate different, more stable behavior across all iodine insert configurations, with a standard deviation to mean ratio of less than 1%. This observed behavior was independent of the area size used for measurement. A minor radiation dose dependency of the electron density values (below 0.5%) was observed. This resulted in consistent (iodine independent) PET quantification behavior, which persisted even at the lowest radiation dose levels tested in our experiment, that is, 25% of the radiation dose utilized for CT acquisition in the clinical PET/CT protocol.
Conclusions:
Utilization of DECT-generated electron density estimations for attenuation correction benefit PET quantification consistency in the presence of iodine and at nominal and low DECT radiation exposure levels. The ability to correctly account for iodinated contrast media in PET acquisitions will allow the development of new clinical applications that rely on the quantitative capabilities of spectral CT technologies and modern PET systems.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...