Standardization and harmonization in hematology: Instrument alignment, quality control materials, and commutability
Matteo Vidali1, Anna Carobene2, Sara Apassiti Esposito3
1Clinical Chemistry Unit, Maggiore della Carità Hospital, Novara, Italy.
Hematology analyzer (HA) alignment can be reliably assessed using quality control (QC) materials, ensuring analytical quality in laboratory networks. However, careful attention is needed for specific white blood cell parameters due to material commutability issues.
Area of Science:
- Clinical Laboratory Science
- Hematology
- Analytical Chemistry
Background:
- Hub and spoke laboratory networks require robust hematology analyzer (HA) alignment for consistent analytical quality.
- Increased numbers of HAs necessitate reliable methods for ensuring instrument comparability.
Purpose of the Study:
- To evaluate a protocol for aligning hematology analyzers (HAs) using quality control (QC) material instead of fresh whole blood.
- To assess the commutability of QC materials for various hematology parameters.
Main Methods:
- Five HAs were aligned using a SIBioC protocol, comparing human samples (HS) and QC materials.
- Commutability was verified per the IFCC protocol, with maximum bias derived from biological variation data.
- Evaluated parameters included red blood cell (RBC), white blood cell (WBC), and platelet (PLT) series.
Main Results:
- Complete alignment was observed for most parameters using both HS and QC materials.
- Partial misalignments or inconclusive results occurred for specific parameters like MCV, MO, EO, BA, and IG.
- Quality control (QC) material showed non-commutability for LY, MO, and BA parameters.
Conclusions:
- Hematology analyzer (HA) alignment for main cell populations can be effectively verified using both QC and HS materials.
- Critical evaluation is needed for certain white blood cell parameters (EO, BA, IG) due to material limitations.
- Careful selection and validation of QC materials are essential for accurate HA alignment.
More Related Videos
06:22Standardization of Transfer across Labs between Flow Cytometers for Detection of Lymphocytes in Japanese Encephalitis Vaccinated Children
Published on: February 10, 2023
08:17A Semi-automated Approach to Preparing Antibody Cocktails for Immunophenotypic Analysis of Human Peripheral Blood
Published on: February 8, 2016
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Standard Solutions
Data Validation
Key parameters for method validation include: