Standardization and harmonization in hematology: Instrument alignment, quality control materials, and commutability
Matteo Vidali1, Anna Carobene2, Sara Apassiti Esposito3
1Clinical Chemistry Unit, Maggiore della Carità Hospital, Novara, Italy.
Introduction:
In the hub and spoke laboratory network, the number of hematology analyzers (HAs) within each core center has increased, and the control of HAs alignment is becoming necessary requirement to ensure analytical quality. In this scenario, HA alignment can be assessed by analyzing the same control material used for internal quality control on multiple HAs, assuming its commutability. The aim of the study was to verify the applicability of a protocol for the alignment of HAs based on control material rather than on fresh whole-blood samples.
Methods:
The alignment of five HAs was evaluated for red (RBC, Hb, MCV, RET), white (WBC, NE, LY, MO, EO, BA, IG), and platelet (PLT) series parameters, following a protocol by SIBioC, using human sample (HS) and quality control material (QC), after the verification of commutability, according to the IFCC protocol. Maximum bias was derived from biological variation data.
Results:
A complete alignment between instruments was confirmed for the majority of the parameters investigated both for HS and QC material. Partial misalignments or inconcludent results were instead evident for MCV, MO, EO, BA, and IG. Interestingly, QC material was found to be not commutable for LY, MO, and BA.
Conclusion:
The alignment of hematologic analyzers for main cell population parameters may be verified with both QC and HS, displaying consistent results and interpretation. The evaluation for some white series parameters (EO, BA, and IG) is critical, and particular attention must be paid to the values of the material used for the alignment.
More Related Videos
06:22Standardization of Transfer across Labs between Flow Cytometers for Detection of Lymphocytes in Japanese Encephalitis Vaccinated Children
Published on: February 10, 2023
08:17A Semi-automated Approach to Preparing Antibody Cocktails for Immunophenotypic Analysis of Human Peripheral Blood
Published on: February 8, 2016
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Standard Solutions
Data Validation
Key parameters for method validation include: