Evaluation of different lead types and implantation techniques in pediatric populations with permanent pacemakers:

Yakup Ergul1, Bekir Yukcu1, Erkut Ozturk1

  • 1Department of Pediatric Cardiology/Electrophysiology, Istanbul Mehmet Akif Ersoy Cardiovascular Research and Training Hospital, Saglik Bilimleri University, Istanbul, Turkey.

Insights

Permanent pacemaker lead failure is a concern in children. This study found that while lead survival rates varied by type, all showed satisfactory early to midterm performance, with transvenous pacing leads demonstrating the best long-term survival.

Area of Science:

  • Pediatric Cardiology
  • Biomedical Engineering
  • Cardiac Electrophysiology

Background:

  • Permanent pacemaker (PM) implantation in children presents challenges, particularly concerning lead performance during follow-up.
  • Lead failure is a significant issue in pediatric PM systems, necessitating evaluation of different implantation techniques and lead types.

Purpose of the Study:

  • To assess the impact of various lead types and implantation techniques on the early and midterm performance of permanent pacemaker systems in pediatric patients.
  • To compare lead electrical performance, failure rates, and survival outcomes across different lead categories.

Main Methods:

  • Retrospective analysis of 323 permanent pacemaker lead implantations in 167 pediatric patients over a 10-year period.
  • Patients were categorized into epicardial pacing (EP), transvenous pacing (TP), and Select Secure (SS) lead groups.
  • Evaluation included demographic data, pacing indications, lead electrical parameters, lead failures, complications, and survival rates.

Main Results:

  • No significant differences were observed in initial or follow-up lead electrical performance (capture, threshold, sensing, impedance) between lead groups.
  • Twenty lead failures occurred during a median follow-up of 3.3 years, with no statistically significant difference in failure rates among EP, TP, and SS leads.
  • Five-year lead survival was 98% for TP, 95% for EP, and 90% for SS; 10-year survival was 90% for TP, 70% for EP, and 70% for SS.

Conclusions:

  • Despite technological advancements, pacemaker lead failure remains a critical concern in pediatric patients.
  • Early to midterm lead survival rates for epicardial, transvenous, and Select Secure leads were satisfactory.
  • Transvenous pacing leads exhibited superior long-term survival rates compared to epicardial and Select Secure leads in this pediatric cohort.
Abstract