Preparing an Isotopically Pure 229Th Ion Beam for Studies of 229mTh
Published on: May 3, 2019
Determination of 226Ra in Urine Using Triple Quadrupole Inductively Coupled Plasma Mass Spectrometry
Ge Xiao1, Yongzhong Liu2, Robert L Jones1
1Inorganic and Radiation Analytical Toxicology Branch, Centers for Disease Control and Prevention, 4770 Buford Highway, Mail Stop S110-5, Atlanta, GA 30341, USA.
Abstract:
Measuring 226Ra in urine at low levels is critical for both biomonitoring and radiological emergency response. Here we report a new analytical method to quantify 226Ra, as developed and validated by a simple dilute-and-shoot procedure, followed by Inductively Coupled Plasma-triple quadrupole-mass spectrometry detection using 'No Gas MS-MS' mode. The method provides rapid and accurate results for 226Ra with a limit of detection (LOD) down to 0.007 ng/l (0.26 Bq/l). This LOD is well below the recommended action levels for 226Ra detection in children and pregnant women (C/P) set by the Clinical Decision Guide (NCRP Report #161). Results for 226Ra obtained by this method are within ±7.0% of the target values of standard reference materials spiked in the urine.