Subject validation of reusable N95 stop-gap filtering facepiece respirators in COVID-19 pandemic
Related Experiment Video
Updated: Nov 30, 2025

Determining Viral Disinfection Efficacy of Hot Water Laundering
Published on: June 21, 2022
William C K Ng1,2,3, Arnaud Romeo Mbadjeu Hondjeu1, Andrew Syrett1
1Department of Anesthesiology and Pain Management, Toronto General Hospital, University Health Network, Toronto, Ontario, Canada.
Introduction:
The COVID-19 pandemic has unveiled widespread shortages of personal protective equipment including N95 respirators. Several centers are developing reusable stop-gap respirators as alternatives to disposable N95 respirators during public health emergencies, using techniques such as 3D-printing, silicone moulding and plastic extrusion. Effective sealing of the mask, combined with respiratory filters should achieve 95% or greater filtration of particles less than 1um. Quantitative fit-testing (QNFT) data from these stop-gap devices has not been published to date. Our team developed one such device, the "SSM", and evaluated it using QNFT.
Methods:
Device prototypes were iteratively evaluated for comfort, breathability and communication, by team members wearing them for 15-30min. The fit and seal were assessed by positive and negative pressure user seal checks. The final design was then formally tested by QNFT, according to CSA standard Z94.4-18 in 40 volunteer healthcare providers. An overall fit-factor >100 is the passing threshold. Volunteers were also tested by QNFT on disposable N95 masks which had passed qualitative fit testing (QLFT) by institutional Occupational Health and Safety Department.
Results:
The SSM scored 3.5/5 and 4/5 for comfort and breathability. The median overall harmonic mean fit-factors of disposable N95 and SSM were 137.9 and 6316.7 respectively. SSM scored significantly higher than disposable respirators in fit-test runs and overall fit-factors (p <0.0001). Overall passing rates in disposable and SSM respirators on QNFT were 65% and 100%. During dynamic runs, passing rates in disposable and SSM respirators were 68.1% and 99.4%; harmonic means were 73.7 and 1643.
Conclusions:
We present the design and validation of a reusable N95 stop-gap filtering facepiece respirator that can match existent commercial respirators. This sets a precedence for adoption of novel stop-gap N95 respirators in emergency situations.
Related Concept Videos
PPE Use in Healthcare Settings I: Donning
PPE Use in Healthcare Settings II: Doffing
Personal Protective Equipment
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...