Diagnostic Reference Levels for Paediatric Head Computed Tomography in Morocco: A Nationwide Survey

M Benmessaoud1, A Dadouch1, M Talbi2

  • 1Faculty of Sciences Dhar EL Mahraz, Sidi Mohamed Ben Abdellah University, B.P. 30000, Fez, Morocco.

Insights

This study established the first Moroccan diagnostic reference levels (DRLs) for paediatric head computed tomography (CT), providing crucial data for radiation dose optimization in children. The findings aim to guide future radiation protection practices in medical imaging across Morocco.

Area of Science:

  • Radiology
  • Medical Physics
  • Paediatric Imaging

Background:

  • Diagnostic Reference Levels (DRLs) are essential for optimizing radiation doses in medical imaging.
  • Paediatric head CT scans require specific DRLs due to children's increased radiosensitivity.
  • Establishing national DRLs is a key component of radiation protection programs.

Purpose of the Study:

  • To determine the initial Diagnostic Reference Levels (DRLs) for paediatric head computed tomography (CT) in Morocco.
  • To assess the effective radiation doses administered during paediatric head CT examinations in Moroccan hospitals.
  • To compare Moroccan DRLs with international standards for paediatric head CT.

Main Methods:

  • Retrospective data collection from 1007 paediatric patients undergoing head CT at Moroccan university children's hospitals.
  • Classification of patients into four age groups: <1, 1-5, 5-10, and 10-15 years.
  • Determination of DRLs using the 75th percentile of CT Dose Index (CTDI) and Dose Length Product (DLP) distributions.

Main Results:

  • Proposed DRLs (75th percentile) for CTDI ranged from 26.98 to 38.20 mGy across age groups.
  • Proposed DRLs (75th percentile) for DLP ranged from 461.64 to 705.98 mGy.cm across age groups.
  • Estimated effective doses varied from 1.79 to 3.6 mSv, decreasing with age.

Conclusions:

  • This study presents the first Moroccan DRLs for paediatric head CT, establishing a benchmark for radiation dose management.
  • The established DRLs provide a basis for comparison with international data and highlight the need for ongoing radiation protection efforts.
  • This initiative serves as a foundation for extending DRL establishment to other imaging procedures and modalities in Morocco.