Related Experiment Video
Updated: Nov 30, 2025

Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
Premature failure of a screw-retained ceramic single crown: A clinical report and fractographic analysis
Daniel Juica1, Natalia Marcus2, Fernando Grandon2
1Graduate student, Postgraduate Program in Oral and Maxillofacial Implantology, Faculty of Dentistry, University of Concepción, Concepción, Chile.
Abstract:
Screw-retained implant-supported ceramic restorations have shown increased rates of technical complications compared with their cemented counterparts, including fracture and chipping of the ceramic structures. The present clinical report identified the causes leading to the catastrophic failure of a screw-retained lithium disilicate veneered crown cemented to a zirconia abutment with a titanium base by using a systematic fractographic approach. A combination of occlusal overloading, a deficient design and inadequate material selection was identified as being responsible for the fracture. These findings highlight the importance of a thorough analysis of the anatomic conditions and loading scenario of screw-retained implant-supported restorations.

