Related Experiment Video
Updated: Nov 30, 2025

Patterning of Microorganisms and Microparticles through Sequential Capillarity-assisted Assembly
Published on: November 4, 2021
Characterization of various geometric arrangements of "air-assisted" flow gating interfaces for capillary
1Faculty of Science, Department of Analytical Chemistry, Charles University, Prague, Czechia.
Abstract:
For connecting flow-through analytical methods with capillary electrophoresis, a chip working in the air-assisted flow gating interface regime is cast from poly(dimethylsiloxane). In the injection space, the exit from the delivery capillary is placed close to the entrance to the separation capillary. Prior to injecting the sample into the separation capillary, the background electrolyte is forced out of the injection space by a stream of air. In the empty space, a drop of the sample with a volume of <100 nL is formed between the exit from the delivery capillary and the entrance into the separation capillary, from which the sample is injected hydrodynamically into the separation capillary. After injection, the injection space is filled with BGE, and the separation can be begun. Three geometric variants for the mutual geometric arrangement of the delivery and separation capillaries were tested: the delivery capillary is placed perpendicular to the separation capillary, from either above or below, or the capillaries are placed axially, that is, directly opposite one another. All of the variants are equivalent from the analytical and separation efficiency viewpoints. The repeatability expressed by RSD is up to 5%. The tested flow gating interface variants are also suitable for continuous and discontinuous sampling at flow rates of the order of units of μL/min. The developed instrument for sequential electrophoretic analysis operates fully automatically and is suitable for rapid sequential monitoring of dynamic processes.
More Related Videos
Related Concept Videos
Capillary Electrophoresis: Instrumentation
Electrophoresis: Overview
There...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...

