Related Experiment Video
Updated: Nov 30, 2025

A Novel Inhalation Mask System to Deliver High Concentrations of Nitric Oxide Gas in Spontaneously Breathing Subjects
Published on: May 4, 2021
Design and construction of a simplified, gas-driven, pressure-controlled emergency ventilator
R Szlosarek1, R Teichert1, A Wetzel1
1Technische Universität Bergakademie Freiberg, Institute for Machine Elements, Design and Manufacturing, Freiberg, Germany.
Introduction:
Due to the COVID-19 crisis or any other mass casualty situation it might be necessary to give artificial ventilation to many affected patients. Contrarily, the worldwide availability of emergency ventilators is still a shortage, especially in developing countries.
Methods:
Modes of artificial ventilation were compared and the most safe, easy to use, and lung protecting principle was optimized to fit all requirements of both emergency ventilation and cost-effective mass production.
Results:
The presented research results describe a simplified device for a pressure-controlled ventilation which works without electricity according to a known principle. Just pressurized gas and a patient connection is required. The device enables the control of basic ventilator parameters such as peak inspiratory pressure, positive end-expiratory pressure and the ventilation frequency. Further, the device is semiadaptive to the patient's lung stiffness and automatically maintains minute volume through frequency adjustment. The machine can be manufactured by turning, milling and drilling and needs purchased components with costs less than 100 USD. A sterilization and thus a reuse is possible.
Discussion:
The presented development does not describe a ready-to-purchase ventilator, it rather outlines a refined working principle for emergency ventilation and its easiest methods of production with a minimum of requirements. The presented research aims on providing an open-source guideline for production of an emergency ventilator using worldwide available methods and thus should inspire local researchers to do a reverse engineering and eventually to put it into operation following country-specific regulations. For long-term ventilation exceeding emergency purposes, a monitoring of alarms for disconnection and violation of desired ventilator parameters should be established. The ventilator is limited to a fixed ratio between PIP and PEEP. Moreover, the ventilation frequency depends on two parameters, which needs some training. Nevertheless, the ventilator provides basic features to enable an emergency ventilation with minimal prerequisites.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cardiopulmonary Resuscitation II: ACLS Airway Management