Swept Source Optical Coherence Tomography Assessment of Bursa Premacularis in Healthy Eyes
Paolo Carpineto1, Rossella D'Aloisio1, Daniele Guarini1
1Ophthalmology Clinic, Department of Medicine and Science of Ageing, University "G. d'Annunzio" Chieti-Pescara, Chieti 66100, Italy.
Purpose:
To describe the reliability and the limits of bursa premacularis (BPM) evaluation using a swept source optical coherence tomography (SS-OCT) device with enface and 16 mm-high definition (HD) longitudinal scans.
Methods:
60 eyes of 60 subjects were enrolled and imaged with SS-OCT system (PLEX Elite 9000, Carl Zeiss Meditec Inc., Dublin, CA, USA). BPM area was measured using enface scans imported to ImageJ. HD horizontal and vertical longitudinal scans centered at the fovea were used to detect width (W) and central thickness (CT) of BPM at baseline (T0) and after 30 minutes (T30) performed by two different observers. An enhanced vitreous visualization software provided by the manufacturer of the device was used to highlight vitreous structures.
Results:
BPM was identifiable in 100% of eyes using both horizontal and vertical longitudinal scans. On horizontal scan, BPM was not entirely measurable in 21.7% and in 18.3% of cases at T0 and T30, respectively. On vertical scan, BPM was not entirely measurable in 75.0% and in 81.7% at T0 and T30, respectively. No statistically significant differences were found between the two different time measurements with an intraclass correlation coefficient above 70%. Median BPM area was 26.9 (Q1-Q3: 19.5-40.5) mm2. In en face imaging, the most frequent BPM shape was the boat one.
Conclusion:
SS-OCT is a reliable tool for a detailed quantification and mapping of BPM, and it is able to add useful details about the morphological BPM features in youth population. However, the enhanced visualization of the vitreous structures is still a challenge, also with the most forefront devices.
More Related Videos
03:35Determining Gender-Based Differences in Retinal and Choroidal Thickness in Underweight Individuals via Swept-Source Optical Coherence Tomography
Published on: December 1, 2023
06:19In Vivo Imaging of Cx3cr1gfp/gfp Reporter Mice with Spectral-domain Optical Coherence Tomography and Scanning Laser Ophthalmoscopy
Published on: November 11, 2017
Related Concept Videos
Glaucoma: Overview
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...