Related Experiment Video
Updated: Nov 29, 2025

The Transition to an Anterior-Based Muscle Sparing Approach Improves Early Postoperative Function but is Associated with a Learning Curve
Published on: September 7, 2022
Protection of the Femoral Trunnion During Acetabular Revision Hip Surgery-A Novel Technique
Abhijit C Kawalkar1, Nick Vertzyas2, Michael J Neil3
1Orthopaedic and Joint Replacement Surgeon, Orange City Hospital and Research Institute, Nagpur, India.
Abstract:
Revision of a failed acetabular component in the presence of an intact femoral prosthesis comprises approximately 20% of revision total hip replacements performed by the senior author. During this procedure, the femoral trunnion is at risk of damage because of scratching from retractors, reamers, or other instruments. Scratching can result in mechanically assisted crevice corrosion or adverse local-tissue reaction or trunnionosis leading to implant failure. We have used a variety of techniques for protection of the femoral taper over the last 30 years. We describe our current method of taper protection, using a cut segment of #5 silicon laryngeal airway, which is a simple and effective method.