Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
Evaluation of analytical performance of Alinity i system on 31 measurands
Jong Do Seo1, Da Young Song1,2, Youngwon Nam3
1Department of Laboratory Medicine, Seoul National University Hospital, Seoul, South Korea.
Introduction:
Accurate, precise and reliable laboratory test results play a critical role in medical decision making. To satisfy the increasing needs in clinical laboratory tests, the analyzers have been advanced. In this study, authors aimed to evaluate the analytical performance of the Alinity i system (Abbott Laboratories, IL, USA) for diverse analytes measured by using immunoassay principle.
Materials And Methods:
Analytical performance of recently launched Alinity i system has been evaluated for 31 assays in aspects of precision, linearity and analytical measurement range, correlation with the Architect i2000sr system (Abbott Laboratories), carry-over, and reference interval validation in accordance with CLSI guidelines.
Results:
The within-laboratory CVs of the analytes tested in the study ranged between 1.00 and 7.84%, which met vendor claimed value in precision. In linearity test, most assays satisfied acceptable linearity criteria, best-fit first order regression or polynomial regression with nonlinearity smaller than ±10%, compared with linear regression. The recovery of each analyte distributed from 90.1 to 109.7%. The coefficient of determination (R2) for each test was larger than 0.95 except for folate when compared to the results obtained from existing routine analyzer and statistically or clinically equivalent. The carry-over rates were acceptable, and reference intervals were validated.
Conclusion:
Through this study, acceptable analytical performance of novel Alinity i system has been verified. It is expected to readily replace existing instrument and to be an option for laboratories considering introduction of automated immunoassay analyzer.
Related Concept Videos
Development of Analytical Methods
Data Validation
Key parameters for method validation include:
Quality Assurance
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Mass Analyzers: Overview
Quantitative Analysis
In quantitative analysis, two key measurements are made: the sample quantity and a property proportional to the amount of the analyte (the substance being analyzed). This forms the basis of the...