Related Experiment Video
Updated: Nov 29, 2025

A Detailed Protocol for Perspiration Monitoring Using a Novel, Small, Wireless Device
Published on: November 24, 2016
Evaluation of a wearable wireless device with artificial intelligence, iThermonitor WT705, for continuous temperature
Yuwei Liu1,2, Changqing Liu3, Min Gao1
1Department of Biliary Surgery, West China Hospital of Sichuan University, Chengdu, China.
Objectives:
To evaluate a new-generation, non-invasive, wireless axillary thermometer with artificial intelligence, iThermonitor (WT705, Raiing Medical, Beijing, China), and to ascertain its feasibility for perioperative continuous body temperature monitoring in surgical patients.
Setting:
Departments of Biliary Surgery and Operating Room and the post-anaesthesia care unit of a university teaching hospital in Chengdu, China.
Participants:
A total of 526 adult surgical patients were consecutively enrolled.
Design:
This was a prospective observational study. Axillary temperatures were continuously recorded with iThermonitor throughout the whole perioperative period. The temperatures of the contralateral armpit were measured with mercury thermometers at 8:00, 12:00, 16:00 and 20:00 every day and were used as references.
Outcome Measures:
The outcomes were the accuracy and precision of the temperatures measured with iThermonitor, the validity to detect fever and the feasibility of continuous wear. Pairs of temperatures were evaluated with Student's t-test, Pearson's correlation and repeated-measures Bland-Altman plot.
Results:
A total of 3621 pairs of body temperatures were obtained. The temperatures measured with iThermonitor agreed with those measured with the mercury thermometers overall, with a mean difference of 0.03°C±0.35°C and a moderate correlation (r=0.755, p<0.001). The 95% limits of agreement (LoA) ranged from -0.63°C to 0.73°C, with 5.11% of the differences outside the 95% LoA. The intraclass correlation coefficient was 0.753. Continuous temperature monitoring captured more fevers than intermittent observation (117/526 vs 91/526, p<0.001), detected fever up to 4.35 hours earlier, and captured a higher peak temperature (0.29°C±0.27°C, 95% CI: 0.26-0.31). All subjects felt that wearing iThermonitor was more or less comfortable and did not affect their daily activities.
Conclusions:
iThermonitor is promising for continuous remote temperature monitoring in surgical patients. However, further developments are still needed to improve the precision of this device, especially for temperature detection in underweight patients and those with lower body temperature.
Trial Registration Number:
ChiCTR1900024549; Results (registered on 5 July 2019).
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...