Direct Pressure Monitoring Accurately Predicts Pulmonary Vein Occlusion During Cryoballoon Ablation
Published on: February 26, 2013
Occlusion tool software for pulmonary vein occlusion verification in atrial fibrillation cryoballoon ablation
Filippo Maria Cauti1, Francesco Solimene2, Giuseppe Stabile2
1Arrhythmology Unit, Ospedale San Giovanni Calibita, Fatebenefratelli Isola Tiberina, Rome, Italy.
Background:
Optimal pulmonary vein (PV) occlusion, usually verified with selective contrast injection, is mandatory to obtain an effective PV isolation during cryoballoon (CB) ablation.
Aim Of The Study:
The purpose of this study was to verify the feasibility and the accuracy of a new dielectric sensing system in assessing PV occlusion during CB ablation in patients with atrial fibrillation (AF).
Methods:
We enrolled 28 consecutive patients with paroxysmal or persistent AF. After transseptal access, a detailed image reconstruction of left atrium and PVs was achieved with an octapolar or decapolar mapping catheter (Achieve catheter, Medtronic Inc.) and KODEX-EPD system (EPD Solutions, a Philips company). The degree of PV occlusion with the inflated Arctic Front Advance Cryoballoon (Medtronic Inc.) was verified using the new "occlusion tool" software module (EPD Solutions, a Philips company) and compared to an angiogram obtained with contrast medium injection in each PV.
Results:
A total of 105 PV CB occlusion were tested. The new occlusion tool software module showed a 91% sensitivity and 76% specificity in assessing a complete PV occlusion verified with contrast medium injection. The positive predictive value was 80%, and the negative predictive value was 88.6%. Mean procedure time was 81 ± 17 minutes. Mean fluoroscopy time was 6 ± 2 minutes. No 30-day complications were observed.
Conclusion:
The new dielectric imaging system was able to assess the degree of PV occlusion during a CB ablation with good sensitivity and specificity.
More Related Videos
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
28:13Catheter Ablation in Combination With Left Atrial Appendage Closure for Atrial Fibrillation
Published on: February 26, 2013
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization