Cardiorespiratory monitoring in the delivery room using transcutaneous electromyography
Ruud W van Leuteren1,2, Eline Kho3,4, Cornelia G de Waal3
1Department of Neonatology, Emma Children's Hospital, Amsterdam UMC, University of Amsterdam, Amsterdam, North-Holland, Netherlands r.w.vanleuteren@amsterdamumc.nl.
Objective:
To assess feasibility of transcutaneous electromyography of the diaphragm (dEMG) as a monitoring tool for vital signs and diaphragm activity in the delivery room (DR).
Design:
Prospective observational study.
Setting:
Delivery room.
Patients:
Newborn infants requiring respiratory stabilisation after birth.
Interventions:
In addition to pulse oximetry (PO) and ECG, dEMG was measured with skin electrodes for 30 min after birth.
Outcome Measures:
We assessed signal quality of dEMG and ECG recording, agreement between heart rate (HR) measured by dEMG and ECG or PO, time between sensor application and first HR read-out and agreement between respiratory rate (RR) measured with dEMG and ECG, compared with airway flow. Furthermore, we analysed peak, tonic and amplitude diaphragmatic activity from the dEMG-based respiratory waveform.
Results:
Thirty-three infants (gestational age: 31.7±2.8 weeks, birth weight: 1525±661 g) were included.18%±14% and 22%±21% of dEMG and ECG data showed poor quality, respectively. Monitoring HR with dEMG was fast (median 10 (IQR 10-11) s) and accurate (intraclass correlation coefficient (ICC) 0.92 and 0.82 compared with ECG and PO, respectively). RR monitoring with dEMG showed moderate (ICC 0.49) and ECG low (ICC 0.25) agreement with airway flow. Diaphragm activity started high with a decreasing trend in the first 15 min and subsequent stabilisation.
Conclusion:
Monitoring vital signs with dEMG in the DR is feasible and fast. Diaphragm activity can be detected and described with dEMG, making dEMG promising for future DR studies.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...
Holter Monitor: 24-Hour Monitoring
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...