Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements
Published on: September 7, 2019
Disregarded Measurement Uncertainty Contributions and Their Magnitude in Measuring Plasma Glucose
Astrid Petersmann1,2, Rainer Macdonald3, Matthias Nauck1,4
1Institute of Clinical Chemistry and Laboratory Medicine, University Medicine Greifswald, Greifswald, Germany.
Background:
Each measurement is subject to measurement uncertainty (MU). Consequently, each measurement of plasma glucose concentration used for diagnosis and monitoring of diabetes mellitus (DM) is affected. Although concepts and methods of MU are well established in many fields of science and technology, they are presently only incompletely implemented by medical laboratories, neglecting MU of target values of internal quality control (IQC) materials.
Methods:
An empirical and practical approach for the estimation of MU based on the analysis of routine IQC using control samples with assigned target values is presented. Its feasibility is demonstrated exemplarily by analyzing IQC data from one year obtained for glucose employing the hexokinase method with IQC of two different concentrations.
Results:
Combined relative extended (k = 2) MU comprising bias, coefficient of variation (CV), and MU of the target values assigned to control materials were about 9% with a lower (~ 56 mg/dL; ~3.1 mmol/L) and 8% with a higher (~ 346 mg/dL; ~19.2 mmol/L) concentration sample, analyzing IQC of one year from three different devices.
Conclusions:
Estimation of MU in this study is quite reliable due to the large number of IQC data from one year. The MU of the target values of the commercial control material in this study was considerably larger than other MU contributions, ie, standard deviation and bias. In the future, the contribution of MU of commercial IQC should be addressed more carefully and technologies to measure glucose should be geared toward smaller MU possible, as needed, especially for glucose concentration measurements in diagnosis and management of DM.
Insights
Measurement uncertainty (MU) affects diabetes mellitus (DM) diagnosis. This study presents a practical method to estimate MU for plasma glucose, finding target value uncertainty in quality control materials is a significant factor.
Area of Science:
- Clinical Chemistry
- Metrology
- Laboratory Medicine
Background:
- All measurements, including plasma glucose, have measurement uncertainty (MU).
- MU impacts diabetes mellitus (DM) diagnosis and monitoring.
- Medical labs often neglect MU of internal quality control (IQC) target values.
Purpose of the Study:
- To present a practical method for estimating MU in routine IQC.
- To demonstrate the feasibility of this MU estimation approach.
- To highlight the impact of IQC target value MU on glucose measurements.
Main Methods:
- Empirical estimation of MU using routine IQC data.
- Analysis of one year of glucose IQC data (hexokinase method).
- Inclusion of bias, CV, and IQC target value MU in combined MU.
Main Results:
- Combined relative extended MU was ~9% (low glucose) and ~8% (high glucose).
- MU estimation was based on one year of IQC data from three devices.
- MU of commercial IQC target values was larger than other MU components.
Conclusions:
- The presented MU estimation method is reliable due to extensive IQC data.
- MU of commercial IQC target values is a critical, often overlooked, factor.
- Future efforts should focus on reducing IQC target value MU for accurate glucose measurement in DM management.