Optimizing and applying high-resolution, in-line laboratory phase-contrast X-ray imaging for low-density material
1Center for X-ray Analytics, Department Materials Meet Life, Swiss Federal Laboratories for Material Science and Technology, Empa, Überlandstrasse 129, Dübendorf, Switzerland.
Journal of Microscopy
|November 21, 2020
Summary
Propagation-based phase-contrast X-ray imaging (PBI) offers superior soft sample imaging over traditional methods. This study optimizes lab-based PBI for enhanced contrast and resolution in low-density materials.
Area of Science:
- Materials Science
- Imaging Physics
- Biomedical Engineering
Background:
- Attenuation-based X-ray imaging struggles with low-density, soft samples due to poor contrast.
- Propagation-based phase-contrast X-ray imaging (PBI) offers an alternative by leveraging phase shifts.
- Advancements in micro- and nano-focus X-ray sources facilitate laboratory-based PBI applications.
Purpose of the Study:
- To demonstrate the advantages of lab-based PBI for low-density materials.
- To present a multi-step optimization process for lab-based PBI systems.
- To evaluate the impact of phase retrieval on image quality and resolution.
Main Methods:
- Utilized a lab-based micro- and nano-computed tomography system (Easytom XL Ultra).
- Performed multi-step optimization: hardware selection, imaging geometry, and phase retrieval.
- Conducted a parametric study to optimize the phase retrieval process.
- Analyzed noise correlation and texturing effects from phase retrieval.
Main Results:
- Achieved micron and sub-micron resolution PBI images of selected samples.
- Demonstrated improved image quality and contrast-to-noise ratio using phase images.
- Observed marginal impact of phase retrieval on spatial resolution.
- Identified noise correlation and texturing as artifacts of phase retrieval.
Conclusions:
- Lab-based PBI significantly enhances imaging of low-density materials compared to transmission imaging.
- Optimized PBI provides superior contrast and image quality, enabling detailed structural analysis.
- Phase retrieval is crucial for maximizing PBI benefits, despite potential artifacts.
Keywords:
X-ray micro/nano CTlow-density materialsnoise correlationphase retrievalpropagation-based phase contrastMore Related Videos
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
11.5K
Phase-Contrast Microscopes
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
11.5K
X-ray Diffraction of Biological Samples
4.4K
X-ray diffraction or XRD is an analytical tool that utilizes X-rays to study ordered structures such as crystalline organic and inorganic samples, polycrystalline materials, proteins, carbohydrates, and drugs.
According to Bragg's law, when X-rays strike the sample positioned on a stage, the rays are scattered by the electron clouds around the sample atoms. The X-ray diffraction or scattering is caused by constructive interference of the X-ray waves that reflect off the internal...
According to Bragg's law, when X-rays strike the sample positioned on a stage, the rays are scattered by the electron clouds around the sample atoms. The X-ray diffraction or scattering is caused by constructive interference of the X-ray waves that reflect off the internal...
4.4K