Epidural Intracranial Pressure Measurement in Rats Using a Fiber-optic Pressure Transducer
Published on: April 25, 2012
Real-Time Intraoperative Pressure Monitoring to Avoid Surgically Induced Localized Brain Injury Using a Miniaturized
Rajat Subhra Karmakar1, Jer-Chyi Wang1,2,3,4, Yu-Ting Huang1
1Department of Electronic Engineering, Chang Gung University, Guishan Dist., Taoyuan 33302, Taiwan.
Abstract:
Neurosurgical procedures often cause damage to the brain tissue at the periphery from surgical manipulations. Especially during retraction, a large amount of pressure could be applied on the brain surface, which can damage it, leading to brain herniation, which can be fatal for patients. To resolve this issue, we have developed a pressure sensor that can be used to monitor the applied pressure during surgery for intraoperative care. This device was tested on a rodent model to create a superficial surgically induced damage profile for three different applied pressures (30, 50, and 70 mmHg) and compared to a standard intracranial pressure monitoring system. Magnetic resonance imaging has been performed after surgical procedures to detect the herniation caused by applied pressure. To evaluate the damage to brain cells and tissue rupture, histological analysis was performed using hematoxylin and eosin staining. A scoring system was developed to understand the severity of the surgically induced brain injury, which will help neurosurgeons to limit the pressure to an optimum point without causing damage.