Computed Tomography Angiography of the Aorta-Optimization of Automatic Tube Voltage Selection Settings to Reduce
André Euler1, Tilo Taslimi1, Matthias Eberhard1
1From the Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland.
Objectives:
The aim of this study was to compare the image quality of low-kV protocols with optimized automatic tube voltage selection (ATVS) settings to reduce either radiation dose or contrast medium (CM) with that of a reference protocol for computed tomography angiography (CTA) of the thoracoabdominal aorta.
Materials And Methods:
In this institutional review board-approved, single-center, prospective randomized controlled trial, 126 patients receiving CTA of the aorta were allocated to one of three computed tomography protocols: (A) reference protocol at 120 kVp and standard weight-adapted CM dose; (B) protocol at 90 kVp, reduced radiation and standard CM dose; and (C) protocol at 90 kVp, standard radiation and reduced CM dose. All three protocols were performed on a third-generation dual-source computed tomography scanner using the semimode of the ATVS system. The image-task-dependent optimization settings of the ATVS (slider level) were adjusted to level 11 (high-contrast task) for protocols A and B and level 3 (low-contrast task) for protocol C. Radiation dose parameters were assessed. The contrast-to-noise ratios (CNRs) of protocols B and C were tested for noninferiority compared with A. Subjective image quality was assessed using a 5-point Likert scale.
Results:
Size-specific dose estimate was 34.3% lower for protocol B compared with A (P < 0.0001). Contrast medium was 20.2% lower for protocol C compared with A (P < 0.0001). Mean CNR in B and C was noninferior to protocol A (CNR of 30.2 ± 7, 33.4 ± 6.7, and 30.5 ± 8.9 for protocols A, B, and C, respectively). There was no significant difference in overall subjective image quality among protocols (4.09 ± 0.21, 4.03 ± 0.19, and 4.08 ± 0.17 for protocols A, B, and C, respectively; P = 0.4).
Conclusions:
The slider settings of an ATVS system can be adjusted to optimize either radiation dose or CM at noninferior image quality in low-kV CTA of the aorta. This optimization could be used to extend future ATVS algorithms to take clinical risk factors like kidney function of individual patients into account.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
09:57Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
Published on: January 18, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography