Assessment of population-based input functions for Patlak imaging of whole body dynamic 18F-FDG PET
Mika Naganawa1, Jean-Dominique Gallezot2, Vijay Shah3
1PET Center, Department of Radiology and Biomedical Imaging, Yale University, New Haven, CT, USA. mika.naganawa@yale.edu.
Background:
Arterial blood sampling is the gold standard method to obtain the arterial input function (AIF) for quantification of whole body (WB) dynamic 18F-FDG PET imaging. However, this procedure is invasive and not typically available in clinical environments. As an alternative, we compared AIFs to population-based input functions (PBIFs) using two normalization methods: area under the curve (AUC) and extrapolated initial plasma concentration (CP*(0)). To scale the PBIFs, we tested two methods: (1) the AUC of the image-derived input function (IDIF) and (2) the estimated CP*(0). The aim of this study was to validate IDIF and PBIF for FDG oncological WB PET studies by comparing to the gold standard arterial blood sampling.
Methods:
The Feng 18F-FDG plasma concentration model was applied to estimate AIF parameters (n = 23). AIF normalization used either AUC(0-60 min) or CP*(0), estimated from an exponential fit. CP*(0) is also described as the ratio of the injected dose (ID) to initial distribution volume (iDV). iDV was modeled using the subject height and weight, with coefficients that were estimated in 23 subjects. In 12 oncological patients, we computed IDIF (from the aorta) and PBIFs with scaling by the AUC of the IDIF from 4 time windows (15-45, 30-60, 45-75, 60-90 min) (PBIFAUC) and estimated CP*(0) (PBIFiDV). The IDIF and PBIFs were compared with the gold standard AIF, using AUC values and Patlak Ki values.
Results:
The IDIF underestimated the AIF at early times and overestimated it at later times. Thus, based on the AUC and Ki comparison, 30-60 min was the most accurate time window for PBIFAUC; later time windows for scaling underestimated Ki (- 6 ± 8 to - 13 ± 9%). Correlations of AUC between AIF and IDIF, PBIFAUC(30-60), and PBIFiDV were 0.91, 0.94, and 0.90, respectively. The bias of Ki was - 9 ± 10%, - 1 ± 8%, and 3 ± 9%, respectively.
Conclusions:
Both PBIF scaling methods provided good mean performance with moderate variation. Improved performance can be obtained by refining IDIF methods and by evaluating PBIFs with test-retest data.
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...