Quantification of the Immunosuppressant Tacrolimus on Dried Blood Spots Using LC-MS/MS
Published on: November 8, 2015
Comparison of Two Analytical Methods for Busulfan Therapeutic Drug Monitoring
Simona De Gregori1, Carmine Tinelli2, Federica Manzoni2
1Clinical and Experimental Pharmacokinetics Unit, Fondazione IRCCS Policlinico San Matteo, Viale Camillo Golgi, 19, 27100, Pavia, Italy. s.degregori@smatteo.pv.it.
Background And Objectives:
Busulfan (Bu) is an old drug, but is still well recommended as an alkylating agent during conditioning therapy, before hematopoietic stem cell transplantation. Although its dose administration is standardized and based on patient weight, therapeutic drug monitoring is required in order to maintain its exposure [as area under the concentration-time curve (AUC) from 0 to infinity AUC0-∞] within a narrow therapeutic range and, if necessary, to adjust the dose with as short a lead time as possible. The aim of the study is to evaluate the agreement (as calculated AUC) between a gold standard analytical method and a new one that is faster and easier.
Methods:
We analyzed 221 plasma samples from 37 children (0.25-16 years; 4-62.5 kg) and 11 adults (21-59 years; 45-80 kg), corresponding to 52 AUC values (ng h/mL). The drug exposure was calculated, simultaneously, by two validated analytical methods. The reference method was a high-performance liquid chromatography (HPLC) assay combined with an ultraviolet detector (UV). The test method had a triple quadrupole mass spectrometer (MS) as detector; the clean-up procedures of the samples were different and faster.
Results:
The agreement between the two methods (reference and test) was evaluated in terms of Bu exposure differences based on Lin's concordance correlation coefficient (CCC) and represented by the Bland-Altman plot. The CCC between the AUC of the two methods was excellent (0.868; 95% CI: 0.802-0.935). The precision of the measures (expressed by Pearson's italic "r") was 0.872, and the accuracy (accounted by the bias correction factor) was 0.996.
Conclusions:
We can conclude that the HPLC-MS/MS assay represents a very good alternative to the reference.
More Related Videos
08:54Validated LC-MS/MS Panel for Quantifying 11 Drug-Resistant TB Medications in Small Hair Samples
Published on: May 19, 2020
09:58Spatial Quantification of Drugs in Pulmonary Tuberculosis Lesions by Laser Capture Microdissection Liquid Chromatography Mass Spectrometry LCM-LC/MS
Published on: April 18, 2018
Related Concept Videos
Therapeutic Drug Monitoring: Drug Analysis Methods
Measurement of Bioavailability: Pharmacodynamic Methods
Measurement of Bioavailability: Pharmacokinetic Methods
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
Drug Concentrations: Measurements
Plasma...
Protein-Drug Binding: Determination Methods
Indirect methods involve isolating the bound drug from its free form in biological samples such as blood, serum, or plasma. These techniques aim to measure the percentage of drugs bound to proteins. Equilibrium dialysis is a commonly used method where the free drug concentration at equilibrium is measured by separating the bound...