Actual delivered dose calculation on intra-irradiation cone-beam computed tomography images: a phantom study
Hiraku Iramina1, Mitsuhiro Nakamura1,2, Takashi Mizowaki1
1Department of Radiation Oncology and Image-Applied Therapy, Graduate School of Medicine, Kyoto University, 54 Kawahara-cho, Shogoin, Sakyo-ku, Kyoto 606-8507, Japan.
Abstract:
Cone-beam computed tomography (CBCT) images acquired during volumetric modulated arc therapy (VMAT; ii-CBCT) can be used to calculate actual delivered doses (ADDs). However, such ii-CBCT images are degraded by scattered megavoltage x-rays (MV-scatters). We aimed to evaluate the dose calculation accuracy of the MV-scatter uncorrected or corrected ii-CBCT images acquired during VMAT deliveries. For MV-scatter correction on concurrent kilovoltage projections (P MVkV), projections consisting only of MV-scatters (P MVS) were acquired under the same MV beam parameters and gantry angles and subtracted from P MVkV (P MVScorr). In addition, the projections by kilovoltage beams were acquired for reference (P kV). The corresponding CBCT images were reconstructed using the Feldkamp-Davis-Kress algorithm (CBCTMVkV, CBCTMVScorr, and CBCTkV as reference). A multi-energy phantom with rods of various relative electron densities (REDs) was used to generate a CBCT-number-to-RED conversion table. First, CBCTkV was reconstructed. Then, the mean CBCT-numbers within each rod were extracted, and a reference table was generated. Concurrent kilovoltage imaging with various field sizes was also demonstrated, and CBCTMVkV and CBCTMVScorr were reconstructed. The extracted CBCT-numbers of each ii-CBCT image were converted into REDs using the reference table. Next, the absolute differences of RED between the ii-CBCT image and CBCTkV were calculated. Ten VMAT plans using a 10 MV flattening-filter-free beam were used for concurrent imaging of an anthropomorphic torso phantom. Moreover, an iterative reconstruction algorithm (IRA) was used for CBCTMVScorr. The plans were recalculated for the corresponding CBCTMVkV, CBCTMVScorr, CBCTMVScorr+IRA, and CBCTkV with the reference table. Finally, the doses were evaluated using 3D gamma analysis (1%/1 mm). The median difference ranges between CBCTMVkV/CBCTMVScorr and the reference values were -0.58 to -0.10/-0.03 to 0.00. The median gamma pass rates of the doses on CBCTMVkV, CBCTMVScorr, and CBCTMVScorr+IRA to the rate on CBCTkV were 70.4, 99.5, and 98.2%, respectively. CBCTMVScorr were comparable with CBCTkV for calculating the ADD from VMAT.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography