Rapid Quantification of Plasma Creatinine Using a Novel Kinetic Enzymatic Assay
Angela M Jimenez Valencia1,2, Lindsay A Kryszak1,2, Joshua Goheen1,2
1Laboratory for Integrated Nanodiagnostics, Johns Hopkins University School of Medicine, Baltimore, MD.
Background:
Enzymatic assays are among the most common diagnostic tests performed in the clinical laboratory. Enzymatic substrate analysis is most commonly measured using endpoint methods; however, modulating the reaction kinetics allows fine control of the reaction rate, which can be adjusted based on specific monitoring technologies.
Methods:
We developed and optimized an enzymatic method for measurement of creatinine in plasma, using commonly paired enzymes of creatininase (Crtnnase), creatinase (Crtase), sarcosine oxidase (SOX), ascorbate oxidase (AOX), and horseradish peroxidase (HRP). The novel aspect of the assay is that it is fast and uses SOX as the limiting enzyme. The assay performance was assessed with respect to precision, accuracy, and interferences.
Results:
The intrarun %CV (n = 12) was approximately 5% for each concentration tested, with biases ranging from -3 to -9%. The interrun %CV (n = 39) ranged from 5 to 8%, with biases ranging from -2 to -6%. During the accuracy assessment (n = 127), only 4 samples did not meet the minimum acceptability criteria. Minimal interference was observed, except at low creatinine concentrations with elevated creatine.
Conclusion:
Our novel and versatile enzymatic assay to measure plasma creatinine using kinetic analysis with SOX as the limiting enzyme is rapid (<2 mins), sensitive, and specific and demonstrates excellent concordance with the laboratory standard. We anticipate this rapid kinetic assay to be compatible with emerging technologies in the field of portable diagnostic devices, such as the usage of silicon photonics to monitor biochemical reactions.
Related Concept Videos
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
Renal Clearance
Renal clearance refers to the volume of plasma cleared of a specific substance, such as creatinine, per unit of time. To measure clearance, urine samples are collected over a 24-hour period during each bladder voiding, followed by a single blood sample at the...
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
Factors Affecting Renal Clearance: Renal Impairment
One condition associated with renal failure is uremia. Uremia is characterized by impaired glomerular filtration and fluid accumulation in the body. This condition hinders the renal clearance of drugs, resulting in drug accumulation and potential...
Renal Drug Clearance: Overview
Renal clearance can be calculated using different methods. One approach is to divide the urinary drug excretion rate by the plasma drug concentration. This method directly measures renal clearance, indicating the kidneys' efficiency in...