Excessive z-axis scan coverage in body CT: frequency and causes.
Ozan Yar1, Mehmet Ruhi Onur2, İlkay Sedakat İdilman3
1Department of Radiology, Artvin Public Hospital, 08000, Artvin, Turkey.
Excessive z-axis coverage in body CT scans is common, occurring in 66% of examinations. Factors like patient condition, scan type, and technologist experience influence this over-scanning, leading to unnecessary radiation exposure.
Area of Science:
- Radiology
- Medical Imaging
- Radiation Safety
Background:
- Excessive z-axis coverage in computed tomography (CT) examinations leads to increased radiation dose.
- Understanding the frequency and causes of over-scanning is crucial for optimizing CT protocols.
Purpose of the Study:
- To determine the incidence of excessive z-axis coverage in body CT examinations.
- To identify patient demographics, examination types, and operational factors contributing to over-scanning.
Main Methods:
- Analysis of 2032 body CT examinations from 1531 patients.
- Calculation of over-scanned length in the z-axis by comparing actual and optimal scan lengths.
- Interrogation of factors including patient age, scan indication, time of day, and technologist experience.
Main Results:
- 66% of body CTs exhibited excessive z-axis coverage.
- Caudal over-scanning (48.5%) was more frequent than cranial (18.4%).
- Higher over-scanning rates were observed in older patients, thorax CTs, patients with altered consciousness, and those with misleading scout images.
Conclusions:
- Over-scanning in the z-axis is a frequent issue in routine body CT practice.
- Specific patient and operational factors significantly increase the likelihood of over-scanning.
- Awareness and mitigation of these factors can reduce patient radiation exposure.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
X-ray Imaging
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography