Accuracy and User Performance of a New Blood Glucose Monitoring System
Leslie Klaff1, Pragathi Shelat2, Diana Zondorak2
1Rainier Clinical Research Center, Renton, WA, USA.
Journal of Diabetes Science and Technology
|November 27, 2020
Summary
The CONTOUR®PLUS ELITE blood glucose monitoring system accurately measures blood glucose levels, exceeding ISO standards. This system offers reliable self-monitoring for people with diabetes to manage their condition effectively.
Area of Science:
- Medical Devices
- Diabetes Technology
- Clinical Diagnostics
Background:
- Self-monitoring of blood glucose (BG) is crucial for diabetes management.
- Accurate BG monitoring informs therapy adjustments and glycemic target achievement.
- Evaluating new blood glucose monitoring systems (BGMS) against established standards is essential.
Purpose of the Study:
- To assess the accuracy and user performance (UP) of the CONTOUR®PLUS ELITE BGMS.
- To evaluate the BGMS against International Organization for Standardization (ISO) 15197:2013 criteria and more stringent benchmarks.
- To determine the system's suitability for people with diabetes (PWD).
Main Methods:
- Laboratory Study 1: 100 PWD capillary fingertip blood samples tested with the BGMS.
- Clinical Study 2: 130 PWD BG measurements compared against YSI analyzer (finger, palm, venous samples).
- User performance assessed via questionnaire; test strip lots evaluated.
Main Results:
- 100% accuracy met ISO criteria; 99.8% met more stringent criteria (±10 mg/dL or ±10%).
- Error grid analysis showed 100% of results in Zone A.
- >98% subject and 100% staff performance met ISO criteria; >96% found the system easy to use.
Conclusions:
- The CONTOUR®PLUS ELITE BGMS surpasses minimum ISO 15197:2013 accuracy and UP standards.
- The system also meets more stringent accuracy criteria.
- This BGMS is a valuable tool for managing glycemic control in people with diabetes.
Related Concept Videos
Errors occurring during blood pressure monitoring
1.2K
Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
Several factors...
1.2K
Equipments Used To Measure Blood Pressure
2.8K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
2.8K
Measurement of Blood Pressure
2.2K
Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a...
2.2K