A New Straightforward Method for Lipophilicity logP Measurement using 19F NMR Spectroscopy
Published on: January 30, 2019
Benchmarking different brands of perfluorocarbon liquids
1, Munich, Germany. joachim@dresp.de.
Quality of perfluorocarbon liquids (PFCL) used in eye surgery varies significantly between brands and batches. This highlights the need for clear manufacturer communication regarding purity to ensure patient safety and accurate surgical outcomes.
Area of Science:
- Ophthalmology
- Materials Science
- Analytical Chemistry
Background:
- Perfluorocarbon liquids (PFCL) are crucial intraoperative tools in vitreoretinal surgery.
- Ensuring the analytical quality of PFCL is essential for patient safety and surgical efficacy.
- Variability in PFCL purity can impact surgical outcomes and complicate interpretation of results.
Purpose of the Study:
- To compare the analytical quality characteristics of CE-marked perfluorocarbon liquids (PFCL).
- To assess the purity profiles of perfluorooctane (PFO) and perfluorodecalin (PFD) used in vitreoretinal surgery.
- To identify potential contaminants and packaging leachables affecting PFCL quality.
Main Methods:
- Analysis of 21 perfluorooctane (PFO) and 25 perfluorodecalin (PFD) samples from 8 and 13 brands, respectively.
- Gas chromatography-mass spectrometry (GC/MS) for main product content determination.
- Fluoride selective potentiometry to quantify reactive impurities (H-value) and UV-active substances for packaging integrity assessment.
Main Results:
- Significant variations in purity profiles were observed across PFO and PFD batches and brands.
- PFO batches showed H-values from <10 to 1230 ppm and 1H-PFO from <1 to 376 ppm.
- PFD batches exhibited H-values from <10 to 70 ppm, with leachables detected in both types of PFCL.
Conclusions:
- The analytical quality of intraoperative PFCLs is highly variable, even within the same manufacturer's batches.
- Clearer communication of product purity by manufacturers is essential.
- Standardized certificates of analysis are needed to help surgeons differentiate PFCL effects from impurity-related complications.