Development of PM10 and PM2.5 cyclones for small sampling ports at stationary sources: Numerical and experimental
Jong-Sang Youn1, Sehyun Han2, Jae-Seong Yi2
1Department of Environmental Engineering, The Catholic University of Korea, 43 Jibong-ro, Bucheon-si, Gyeonggi-do, 14662, Republic of Korea.
Abstract:
Air pollution caused by particulate matter (PM) has become a serious issue, and significant research has focused on managing large stationary emission sources, i.e., the primary sources of PM. Currently, the U.S. Environmental Protection Agency (EPA) Method 201A and ISO 23210 are predominantly employed to measure the PM emissions at large stationary sources. Method 201A is designated as a standard test method in Korea, but it is difficult to measure PM10 and PM2.5 simultaneously owing to the size of the full-set cyclone. In large stationary emission sources, the use of a serial connection of PM10 and PM2.5 cyclones is unsuitable for measurements at conventional sampling ports featuring diameters of approximately 100 mm. Therefore, in this study, PM10 and PM2.5 cyclones were developed to replace the cyclones currently used in Method 201A. The developed cyclones featured a cutoff diameter, which was confirmed by numerical and experimental analyses that were close to Method 201A. Moreover, there was an increase in the stiffness of collection efficiency. The hook adaptor, which is a key accessory used in Method 201A, was found to be applicable to the newly developed cyclones. This alternative method will help reduce the measurement time by simultaneously measuring TSP, PM10, and PM2.5 and eliminates the costs of installing or refurbishing additional sampling ports at existing large stationary sources.
Related Concept Videos
Precipitate Formation and Particle Size Control
The obtained precipitate should be either a pure substance of known composition or easily converted to one by a simple process, such as ignition or drying. In addition, the precipitate should be insoluble and easily filterable. In general, filterability...
Sampling Methods: Sample Types
Solid samples include a variety of substances, such as sediments from water bodies, soil, metals, and biological tissues. Two standard methods for extracting sediments from water bodies are grab sampling and piston coring. Grab sampling involves using a device to collect a discrete sediment sample from the bottom of a water body with minimal disturbance. Grab samples do not always represent the entire area due to...