Related Experiment Video
Updated: Nov 28, 2025

Imaging In-Stent Restenosis: An Inexpensive, Reliable, and Rapid Preclinical Model
Published on: September 14, 2009
Healing and early stent coverage after ultrathin strut biodegradable polymer-coated sirolimus-eluting stent
Atul Abhyankar1, Alexandre Abizaid2, Daniel Chamié3
1Department of Cardiology, Shree B. D. Mehta Mahavir Heart Institute, Surat, India.
Aims:
The aim of SiBi study was to evaluate the early vascular healing and neointimal coverage after implantation of ultrathin (60 μm) biodegradable polymer-coated Tetriflex (Sahajanand Medical Technologies Pvt. Ltd., Surat, India) sirolimus-eluting stent (SES) using optical coherence tomography (OCT) at 4 to 6 weeks after implantation.
Methods:
SiBi was a single-center, observational, investigator-initiated study. From January 15, 2018 to April 15, 2018, total 29 consecutive patients who had consented and underwent OCT examination at 4-6 weeks after Tetriflex SES implantation were enrolled. All OCT images were analyzed at an independent core laboratory by analysts who were blinded to patient and procedural information.
Results:
Of 29 patients, four patients were excluded, as those OCT images were technically inadequate for analysis. Therefore, 25 patients were included in final OCT analysis. Average OCT analysis was performed after 35.3 ± 5 days of Tetriflex implantation. Total 14,024 stent struts in 1,520 cross sections were analyzed. Strut tissue coverage was observed in 91.26 ± 5.53% of struts and malapposed struts were seen in 0.89 ± 1.67%. The mean neointimal hyperplasia (NIH) thickness on the covered struts was 50 ± 30 μm.
Conclusion:
A large percentage of struts were found to be covered with thin layer of NIH evenly distributed along the stent length at around 1 month from stent implantation. The results of this pilot study serve as ethical and scientific backbone to conduct an adequately powered clinical trial to evaluate outcomes of short dual-antiplatelet therapy in context of ultrathin strut stent.