Biomechanical Study Comparing Pulvertaft, Double Side-to-Side, and Locking Side-to-Side Tendon Suture Techniques
Jaimy E Koopman1, Caroline A Hundepool1, Liron S Duraku1
1Department of Plastic, Reconstructive Surgery and Hand Surgery, Erasmus Medical Centre, Rotterdam, The Netherlands.
The Journal of Hand Surgery
|November 30, 2020
Summary
The single-sided locking side-to-side (STS) suture technique offers a stronger, less bulky alternative to the Pulvertaft weave for tendon transfers. This new method is particularly effective for extensor tendon repairs.
Area of Science:
- Orthopedic Surgery
- Biomedical Engineering
- Musculoskeletal Research
Background:
- Tendon transfers are crucial surgical procedures for restoring function.
- The Pulvertaft weave is a common technique for tendon reconstruction.
- There is a need for stronger and less bulky alternatives in tendon repair.
Purpose of the Study:
- To evaluate the biomechanical properties of a novel single-sided locking side-to-side (STS) suture technique.
- To compare the single-sided locking STS technique against the Pulvertaft weave and a double-sided nonlocking STS technique.
- To determine the suitability of these techniques for tendon transfers.
Main Methods:
- Utilized 24 human cadaveric extensor digitorum communis and 24 flexor digitorum superficialis tendons.
- Randomly assigned tendons into three groups: Pulvertaft, double-sided nonlocking STS, and single-sided locking STS.
- Measured load to failure and increase in cross-sectional area (bulkiness) using tensile testing.
Main Results:
- Single-sided locking STS reconstructions demonstrated higher load to failure than Pulvertaft in both extensor and flexor tendons.
- No significant difference in load to failure was observed between single-sided locking STS and double-sided nonlocking STS in extensor tendons.
- Pulvertaft reconstructions showed greater bulkiness compared to both single-sided locking and double-sided nonlocking STS techniques.
Conclusions:
- Single-sided locking and double-sided nonlocking STS reconstructions are viable, stronger, and less bulky alternatives to the Pulvertaft technique.
- The single-sided locking STS reconstruction is particularly well-suited for extensor tendon transfers due to comparable strength and reduced bulk.
- These findings support the clinical relevance of STS techniques for improved tendon transfer outcomes.