Obstetric forceps dimensions and the newborn head biometry: Time for an update
Abdul-Qader T Ismail1, Anne Fritze2, Mario Rüdiger2
1New Cross Hospital, The Royal Wolverhampton NHS Trust, Wolverhampton, WV10 0QP, UK.
European Journal of Obstetrics, Gynecology, and Reproductive Biology
|December 1, 2020
Summary
Obstetric forceps design has not changed in 400 years, despite modern labor practices. Optimizing forceps dimensions based on newborn head biometry could reduce neonatal complications from assisted vaginal delivery.
Area of Science:
- Obstetrics
- Medical Device Design
- Neonatal Health
Background:
- Obstetric forceps, designed in the 1600s for severely molded fetal heads during prolonged labor, have dimensions largely unchanged over 400 years.
- Modern obstetric practice aims to avoid protracted labor, making historical forceps design potentially suboptimal.
- Existing forceps dimensions may not align with current newborn head biometry.
Purpose of the Study:
- To re-evaluate and propose optimized dimensions for obstetric forceps.
- To base new dimensions on contemporary newborn head biometry, specifically biparietal and mentovertical diameters.
- To hypothesize that optimized forceps will reduce neonatal complications and improve maternal outcomes.
Main Methods:
- Systematic review of existing literature on forceps design and neonatal outcomes.
- Original measurements of neonatal mentovertical and biparietal diameters using laser scanning technology.
- Correlation analysis between neonatal head biometry and current forceps dimensions.
Main Results:
- Demonstrated a significant discrepancy between historical forceps dimensions and modern neonatal head biometry.
- Identified specific correlations between biparietal diameter and interblade distance, and mentovertical diameter and blade length.
- Established a rationale for redesigning forceps based on these biometric findings.
Conclusions:
- Current obstetric forceps dimensions are outdated and may not be optimal for contemporary newborns.
- Optimizing forceps dimensions based on neonatal head biometry is proposed as a method to reduce delivery complications.
- Further research and clinical trials are warranted to validate the impact of redesigned forceps on neonatal and maternal outcomes.