Related Experiment Video
Updated: Nov 27, 2025

Implantation of Electroencephalogram and Electrocardiogram Telemetry Devices in Neonatal Rabbit Kits
Published on: February 28, 2025
Sudden cardiac arrest in an epicardial paced-dependent child: watch out, it's a pitfall!
Raymond N Haddad1, Sophie Malekzadeh-Milani1, Damien Bonnet1
1M3C-Necker, Hôpital Universitaire Necker-Enfants malades, University de Paris, Paris, France.
Abstract:
Coronary artery compression by epicardial leads is a rare complication in children and can be difficult to identify with potentially lethal outcomes. Herein, we report the case of a previously asymptomatic paced-dependant 5-year-old girl who presented to our institution with resuscitated cardiac arrest. We describe the atypical sequence of clinical findings misleading initial diagnosis. Hardware failure and the commonly occurring lead fracture were incriminated in the mechanism of cardiac arrest, precipitating implantation of a new pacing system while concealing dynamic compression of the left anterior descending coronary artery.
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
Cardiopulmonary Resuscitation IV: Pharmacological Management
Introduction Cardiac Emergencies
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Cardiopulmonary Resuscitation I: Adult