Related Experiment Video
Updated: Nov 27, 2025

A Detailed Protocol for Perspiration Monitoring Using a Novel, Small, Wireless Device
Published on: November 24, 2016
A Capacitive Sweat Rate Sensor for Continuous and Real-Time Monitoring of Sweat Loss
Dong-Hoon Choi1, Mark Gonzales1, Grant B Kitchen1
1Institute for Nanobiotechnology, John Hopkins University, 3400 North Charles Street, Baltimore, Maryland 21218, United States.
Abstract:
Individualized measurement of sweat loss under heat stress is important in assessing physical performance and preventing heat-related illness for athletes or individuals working in extreme environments. The objective of this work was to develop a low-cost and easy-to-fabricate wearable sensor that enables accurate real-time measurement of sweat rate. A capacitive-type sensor was fabricated from two conducting parallel plates, plastic insulating layers, and a central microfluidic channel formed by laser cutting a plastic film. The device has no microfabricated electrodes and is assembled using adhesive tape. Sensor accuracy was validated at different flow rates and confirmed using an equivalent circuit model of the device. On-body measurements demonstrate the feasibility of real-time measurements and show good agreement with values determined from a conventional sweat collection device.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Accessory Structures of the Skin: Sweat Glands
Sweat glands are classified as merocrine glands; that is, the secretions are excreted by exocytosis through a duct without affecting the cells of the gland. There...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...