Related Experiment Video
Updated: Nov 27, 2025

Quantification of Oculomotor Responses and Accommodation Through Instrumentation and Analysis Toolboxes
Published on: March 3, 2023
Development of an Age-corrected Normative Database for Saccadic Vector Optokinetic Perimetry (SVOP)
Andrew J Tatham1, Pam McClean1, Ian C Murray1
1Department of Ophthalmology, University of Edinburgh.
Precis:
Normal age-corrected threshold sensitivity values were determined for a new eye tracking perimeter and compared with standard automated perimetry (SAP).
Purpose:
The purpose of this study was to determine threshold visual field sensitivities in normal subjects performing saccadic vector optokinetic perimetry (SVOP), a new eye tracking perimeter.
Patients And Methods:
A total of 113 healthy participants performed SVOP and SAP in both eyes with the order of testing randomized. The relationship between SAP and SVOP sensitivity was examined using Bland-Altman plots and 95% limits of agreement. The relationship between sensitivity and age was examined by pointwise linear regression and age-corrected normal threshold sensitivities were calculated.
Results:
After excluding unreliable tests, 97 participants with a mean age of 65.9±10.1 years were included. Average SAP mean deviation was -0.87±1.56 dB, SAP sensitivity was 29.20±1.68 dB and SVOP sensitivity was 32.18±1.96 dB. SVOP had a longer test duration (431±110 compared with 307±42 seconds for SAP, P<0.001). On average, the mean sensitivity obtained using SVOP was 2.98 dB higher than average SAP sensitivity, with 95% limits of agreement of -0.11 to 6.15 dB. For each decade older, SAP sensitivity decreased by 0.93 dB (95% confidence interval: 1.21 to 0.64) and SVOP sensitivity decreased by 1.15 dB (95% confidence interval: 1.47 to 0.84).
Conclusions:
The results provide age-corrected normative values for threshold sensitivities from SVOP. Overall, SVOP provided a similar shaped hill of vision as SAP however threshold sensitivities were higher, meaning results are not interchangeable.