Reducing the influence of geometry-induced gradient deformation in liquid chromatographic retention modelling
Tijmen S Bos1, Leon E Niezen2, Mimi J den Uijl2
1Division of Bioanalytical Chemistry, Amsterdam Institute for Molecular and Life Sciences, Vrije Universiteit Amsterdam, De Boelelaan 1085, 1081 HV Amsterdam, The Netherlands; Centre for Analytical Sciences Amsterdam (CASA), The Netherlands.
Abstract:
Rapid optimization of gradient liquid chromatographic (LC) separations often utilizes analyte retention modelling to predict retention times as function of eluent composition. However, due to the dwell volume and technical imperfections, the actual gradient may deviate from the set gradient in a fashion unique to the employed instrument. This makes accurate retention modelling for gradient LC challenging, in particular when very fast separations are pursued. Although gradient deformation has been addressed in method-transfer situations, it is rarely taken into account when reporting analyte retention parameters obtained from gradient LC data, hampering the comparison of data from various sources. In this study, a response-function-based algorithm was developed to determine analyte retention parameters corrected for geometry-induced deformations by specific LC instruments. Out of a number of mathematical distributions investigated as response-functions, the so-called "stable function" was found to describe the formed gradient most accurately. The four parameters describing the model resemble the statistical moments of the distribution and are related to chromatographic parameters, such as dwell volume and flow rate. The instrument-specific response function can then be used to predict the actual shape of any other gradient programmed on that instrument. To incorporate the predicted gradient in the retention modelling of the analytes, the model was extended to facilitate an unlimited number of linear gradient steps to solve the equations numerically. The significance and impact of distinct gradient deformation for fast gradients was demonstrated using three different LC instruments. As a proof of principle, the algorithm and retention parameters obtained on a specific instrument were used to predict the retention times on different instruments. The relative error in the predicted retention times went down from an average of 9.8% and 12.2% on the two other instruments when using only a dwell-volume correction to 2.1% and 6.5%, respectively, when using the proposed algorithm. The corrected retention parameters are less dependent on geometry-induced instrument effects.
Related Concept Videos
Chromatographic Resolution
The effectiveness of separation can be evaluated by determining the level of separation between two neighboring peaks in a chromatogram, which represents the individual components of a sample.
In chromatography,...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Diffusion on Chromatography Columns
Longitudinal diffusion occurs when the solute molecules in the mobile phase diffuse from the more concentrated center of the chromatographic band to the more dilute regions on either side, both towards and against the flow direction. This...
High-Performance Liquid Chromatography: Elution Process
Chromatographic Methods: Terminology
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process: