Related Experiment Video
Updated: Nov 27, 2025

In Vitro Biochemical Assays using Biotin Labels to Study Protein-Nucleic Acid Interactions
Published on: July 17, 2019
Biotin Interference in Point of Care HIV Immunoassay
Mohan Kumar Haleyur Giri Setty1, Sherwin Lee1, Julia Lathrop2
1Laboratory of Molecular Virology, Office of Blood Research and Review, CBER, FDA, Silver Spring, Maryland, USA.
Abstract:
The use of high concentrations of biotin as a dietary supplement to improve hair, skin, and nail quality has increased in the United States over the past few years. High concentrations of biotin have been shown to interfere with some diagnostic assays that use streptavidin-biotin interactions as one of the steps in the assay. The objective of this report is to evaluate potential biotin interference on the analytical and clinical sensitivity of a point of care (POC) antigen-antibody combo HIV-1 assay. We spiked biotin at concentrations ranging from 12.5 to 400 ng/mL into serum and plasma containing HIV-1 subtype B p24 antigen derived from culture supernatant. The p24 antigen was present in the matrices at 30 pg/mL. Fifty microliters of each sample was applied to Alere Determine HIV-1/2 Ag/Ab combo assay strips in duplicate and results were read by eye after 20 to 30 min. Biotin interfered with detection of HIV-1 p24 in serum and plasma. HIV-1 p24 was not detected at 30 pg/mL p24 when biotin was present at 200 ng/mL concentration. Our study demonstrated that elevated levels of biotin in samples may interfere with POC assays. It is important to consider biotin supplements as potential sources of falsely increased or decreased test results, especially in cases wherein supplementation cannot be ruled out.