In Vitro Application of a Wireless Sensor in Flexion-Extension Gap Balance of Unicompartmental Knee Arthroplasty
Published on: May 5, 2023
A Novel Technique for Varus Tibial Cutting for Oxford Unicompartmental Knee Arthroplasty
Takafumi Hiranaka1, Toshikazu Tanaka1, Takaaki Fujishiro1
1Department of Orthopaedic Surgery and Joint Surgery Centre, Takatsuki General Hospital, Takatsuki, Japan.
Abstract:
To reduce the stress on the medial tibial cortex and to decrease the risk of fracture, a varus cut of the tibia appears to be a reasonable alternative to the orthogonal cut by conventional methods. We present a new instrument and procedure, which enables a varus tibial cut for Oxford unicompartmental knee arthroplasty. We used a custom-made, slidable fixator instead of the standard fixator to set the extramedullary rod on the leg. We also made a numeric formula and a chart to arrange the varus cutting angle using the length of the mediolateral shift of the distal end and the longitudinal extension length of the extramedullary tibial rod. A varus cut up to 4.5° can be controlled. This technique is a simple and useful means of obtaining a varus tibial cut for Oxford unicompartmental knee arthroplasty.