Technical Note: Characteristics of a microSilicon X shielded diode detector for photon beam dosimetry
Yuichi Akino1, Keita Okamura2, Indra J Das3
1Oncology Center, Osaka University Hospital, 2-2 (D10), Yamadaoka, Suita, Osaka, 565-0871, Japan.
Purpose:
This study investigated the characteristics of a new shielded diode detector, microSilicon X (model 60022: MSX), for small-field and large-field dosimetry.
Methods:
The percent depth dose (PDD), beam profiles, detector output factor (OFdet ), temperature dependence, dose rate dependence, dose-per-pulse (DPP) dependence, and dose-response linearity of MSX were evaluated in Cyberknife and TrueBeam photon beams and compared with various detectors including microDiamond (PTW model 60019: MD), Sun Nuclear EDGE detector, Photon diode (PTW model 60016: PD), and semiflex ionization chamber (PTW model 31010: IC).
Results:
For field sizes ranging from 50 × 50 mm2 to 400 × 400 mm2 , MSX-measured OFdet values were within 1% of the IC-measured values. For the CyberKnife small fields, the maximum difference between the MSX-measured OFdet and the MD-measured field output factor (Ω) was 4.0%, while the maximum differences were 8.8% and 10.9% for PD and EDGE, respectively. MSX showed a stable response within 0.7% for water temperatures of 5°C to 34°C, while PD and EDGE showed a linear correlation between the water temperature and the response. MSX showed small variations within 0.2% for the dose rate, and PD and EDGE showed logarithmic increases in the response with the dose rate. MSX and MD had smaller DPP dependences than PD and EDGE.
Conclusions:
The characteristics of MSX for measurements of small- and large-field photon beams are favorable. Compared to PD, MSX exhibited significant improvement in the over-response for small fields. The OFdet values measured by MSX were approximately in-between those measured by MD and PD. MSX showed stable responses against water temperature, dose rate, and DPP variations and provided suitable data for a wide range of field sizes. However, careful attention is required for measurements of OFdet for field sizes of <10 × 10 mm2 and PDD for field sizes of ≥200 × 200 mm2 .