Background/Purpose:
Coagulation complications are frequent, unwanted occurrences in extracorporeal membrane oxygenation (ECMO) treatment, possibly influenced by the pump in the ECMO-circuit. We hypothesized that fewer complications would occur with a smaller, heparin-coated ECMO system with a centrifugal pump (CP) than with one with a roller pump (RP) and that after conversion, complication rates would decrease over time.
Methods:
This single-center, retrospective chart study included all first neonatal and pediatric ECMO runs between 2009 and 2015. Differences between groups were assessed with Mann-Whitney U tests and Kruskal-Wallis tests. Determinants of complication rates were evaluated through Poisson regression models. The CP group was divided into three consecutive groups to assess whether complication rates decreased over time.
Results:
The RP group comprised 90 ECMO runs and the CP group 82. Hemorrhagic complication rates were significantly higher with the CP than with the RP, without serious therapeutic consequences, while thrombotic complications rates were unaffected. Intracranial hemorrhage rates and coagulation-related mortality rates were similar. Gained experience with the CP did not improve complication rates or survival over time.
Conclusions:
Although the CP seems safe, it does not seem beneficial over the RP. Further research is warranted on how pump type affects coagulation, taking into account the severity and implications of coagulation complications.
Level Of Evidence:
Level III.
Related Concept Videos
Hemodialysis II: Procedure and Complications
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications
Extrinsic and Intrinsic Pathways of Hemostasis
The Extrinsic Pathway
The extrinsic pathway of coagulation is typically initiated by tissue damage that exposes blood to tissue factor (TF), a protein released by the damaged tissue cells outside the blood vessels—this interaction with TF triggers biochemical reactions involving specific clotting factors. The key player here is Factor VII, which...
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy