Paul M Meaney1, Andrew P Gregory2, Tapani Lahtinen3
1Thayer School of Engineering, Dartmouth College, Hanover, NH 03755 and Chalmers University of Technology, Gothenburg Sweden 412 96.
Abstract:
In the paper, "Investigation of histology region in dielectric measurements of heterogeneous tissues," by Porter and O'Halloran, the authors utilize a flexible phantom in a layered material dielectric property analysis to quantify the effective sensing volume of a coaxial dielectric probe. Ostensibly, this test has been used by others to characterize the region for which percent variation in the material composition in front of the probe corresponds to percent variation in the computed effective dielectric properties. By employing a compressible material, the authors fail to isolate features that are attributable solely to the probe, itself, and inadvertently incorporate confounding characteristics associated with the compressible nature of the material. The net effect is to exaggerate the probe's sensing volume which undermines conclusions drawn from the subsequent tissue dielectric property studies.