Application of large volume injection for sensitive LC-MS/MS analysis of seven artificial sweeteners in surface
1Institute of Inorganic and Analytical Chemistry, University of Münster, Corrensstraße 30, 48149 Münster, Germany.
Abstract:
The combination of large volume injection and mixed-mode chromatography was performed for direct ultra-trace LC-MS/MS analysis of seven artificial sweeteners with varying physicochemical properties in surface water samples.•The injection volume was raised from 10 µL to 500 µL, while the overall analysis time was only increased by ≈5 min compared to the initial method.•Online column head refocusing and concentration of analytes enabled detection in sub-ng L-1 concentration range without elaborate sample preparation steps.•Relative standard deviations <7% despite multiple injection into the loop.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
11:59Extraction and Detection of Geosmin and 2-Methylisoborneol in Water and Fish using High-Capacity Sorptive Extraction Probes and GC-MS
Published on: July 3, 2025