Marjolaine Georges1,2,3,4, Claudio Rabec5, Elise Monin5
1Department of Pulmonary Medicine and Intensive Care Unit, University Hospital, Dijon, France. marjolaine.georges@chu-dijon.fr.
Background:
Noninvasive ventilation (NIV) represents an effective treatment for chronic respiratory failure. However, empirically determined NIV settings may not achieve optimal ventilatory support. Therefore, the efficacy of NIV should be systematically monitored. The minimal recommended monitoring strategy includes clinical assessment, arterial blood gases (ABG) and nocturnal transcutaneous pulsed oxygen saturation (SpO2). Polysomnography is a theoretical gold standard but is not routinely available in many centers. Simple tools such as transcutaneous capnography (TcPCO2) or ventilator built-in software provide reliable informations but their role in NIV monitoring has yet to be defined. The aim of our work was to compare the accuracy of different combinations of tests to assess NIV efficacy.
Methods:
This retrospective comparative study evaluated the efficacy of NIV in consecutive patients through four strategies (A, B, C and D) using four different tools in various combinations. These tools included morning ABG, nocturnal SpO2, TcPCO2 and data provided by built-in software via a dedicated module. Strategy A (ABG + nocturnal SpO2), B (nocturnal SpO2 + TcPCO2) and C (TcPCO2 + builtin software) were compared to strategy D, which combined all four tools (NIV was appropriate if all four tools were normal).
Results:
NIV was appropriate in only 29 of the 100 included patients. Strategy A considered 53 patients as appropriately ventilated. Strategy B considered 48 patients as appropriately ventilated. Strategy C misclassified only 6 patients with daytime hypercapnia.
Conclusion:
Monitoring ABG and nocturnal SpO2 is not enough to assess NIV efficacy. Combining data from ventilator built-in software and TcPCO2 seems to represent the best strategy to detect poor NIV efficacy. Trial registration Institutional Review Board of the Société de Pneumologie de Langue Française (CEPRO 2016 Georges).
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...