Optimization of scatter estimation window setting for quantitative analysis in 111 In-pentetreotide single photon
Mizuho Hishikawa1, Norikazu Matsutomo, Tomoaki Yamamoto
1Graduate School of Health Sciences, Department of Medical Radiological Technology, Faculty of Health Sciences, Kyorin University, 5-4-1 Shimorenjaku Mitaka, Tokyo 181-8612, Japan. tyamamoto@ks.kyorin-u.ac.jp.
Objective:
The aim of this study was to validate the optimal scatter correction method and scatter estimation window setting in terms of image quality and quantitative accuracy for quantitative indium-111 (111In)-pentetreotide SPECT imaging.
Materials And Methods:
We used a positron emission tomography/computed tomography (PET/CT) phantom to validate image quality and quantitative accuracy, and the SPECT images were acquired by the multi-energy window (MEW) method. The scatter estimation was performed using four kinds of energy windows (MEW1, MEW2, MEW3, and MEW4). Scatter correction was also performed using a dual-energy window (DEW) for comparison with MEWs. Image quality was assessed using percent contrast (% contrast) and background variability, and quantitative accuracy was assessed using the mean standardized uptake value (SUVmean) with hot spheres.
Results:
In the quantification, all MEW settings approached the theoretical SUVmean (MEW1, 0.99±0.06; MEW2, 0.99±0.05; MEW3, 1.00±0.08; MEW4, 0.97±0.12) in contrast to DEW (0.88±0.05). The SUVmean value for scatter correction of both photopeaks for a 28 mm sphere showed the smallest difference from the theoretical value.
Conclusion:
The scatter correction method that gave optimal image quality and quantitative accuracy was MEW3 with two 20% energy windows (one over each photopeak) and four adjacent 3% scatter estimation windows (one on each side of the two photopeaks).
More Related Videos
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
08:02Author Spotlight: Enhanced Quantification of Cardiovascular Calcification Progression for Longitudinal Micro PET/CT Studies in Small Research Animals
Published on: November 15, 2024
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...