Comparison of the accuracy of dental implant placement using static and dynamic computer-assisted systems: an in
Miao Zhou1, Hui Zhou2, Shu-Yi Li1
1Department of Oral and Maxillofacial Surgery, Affiliated Stomatology Hospital of Guangzhou Medical University, Guangzhou Key Laboratory of Basic and Applied Research of Oral Regenerative Medicine, Huangshadadao Road 39, 510182, Guangzhou, China.
Objectives:
The study aimed to compare the accuracy of implant placement between static and dynamic computer-assisted systems (CAS) in a partially edentulous mandible model.
Materials & Methods:
A total of 80 implants was placed in mandible models. The implants were placed using either static or dynamic computer-assisted system. Deviations of implant hex, apex and angulation were measured between preoperative planning and postoperative CBCT in planning software.
Results:
The mean deviations of implant hex, apex and angulation in static CAS group were 1.15 ± 0.34 mm, 1.37 ± 0.38 mm and 2.60 ± 1.11 degree, while in dynamic CAS group were 0.40 ± 0.41 mm, 0.34 ± 0.33 mm and 0.97 ± 1.21 degree, respectively. Implant placement with dynamic CAS showed less deviations of shoulder, apex and angulation than with static CAS significantly.
Conclusions:
The implant accuracy using CAS system could be influenced by the guiding technique.