Related Experiment Video
Updated: Nov 26, 2025

Signal Acquisition, Score Interpretation, and Economics of a Non-Invasive Point-of-Care Test for Coronary Artery Disease
Published on: August 9, 2024
Implementing Point-of-Care Troponin Testing in the Emergency Department: Impact on Time to Result
Introduction:
In the emergency department, troponin assays are commonly used and essential in the evaluation of chest pain and diagnosis of acute coronary syndrome. This study was designed to assess the potential impact of implementing point-of-care troponin testing by comparing the time to point-of-care laboratory result and time to conventional laboratory result.
Methods:
The study enrolled 60 ED patients deemed to need a troponin test in the evaluation of low-risk chest pain (HEART score <4 based on history, electrocardiogram, age, risk factors). Point-of-care troponin testing was performed with the same blood sample obtained for a conventional troponin assay. If the provider determined that the patient required 2 troponin tests, the second laboratory draw was used in the data collection. This was to correlate the time of laboratory result to time of disposition.
Results:
Of the 60 subjects enrolled, 2 subjects were excluded because of user errors with the point-of-care testing equipment and 2 others for not meeting inclusion criteria on later review. The median times for the point-of-care troponin and conventional troponin assays were 11:00 minutes (interquartile range 10:00-15:30) and 40:00 minutes (interquartile range 31:30-52:30), respectively; P < 0.001. There were 3 extreme outliers from the conventional troponin assay that significantly skewed the distribution of the mean, making the median the more accurate assessment of the central tendency.
Discussion:
Point-of-care troponin testing provided results in a median time 29 minutes quicker than the conventional troponin assay. This result is statistically significant and has the potential to greatly improve time to disposition in all patients with chest pain requiring a troponin assay.
Related Concept Videos
Acute Coronary Syndrome III: Diagnostic Studies
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Acute Coronary Syndrome IV: Interprofessional Care
Acute Coronary Syndrome I: Introduction
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Chronopharmacokinetics: Circadian Rhythms and Influence on Drug Response
The time of drug administration is an important factor to consider, as it can influence the toxic dose of a drug. For example, a study conducted by Prins et al. in 1997 examined the effects of the timing of...