European experience with a first totally leadless cardiac resynchronization therapy pacemaker system
Adrien Carabelli1, Mariem Jabeur1, Peggy Jacon1
1Arrhythmias Unit, Department of Cardiology, Grenoble University Hospital, CS 10217, 38043 Grenoble Cedex 09, France.
Summary
Totally leadless cardiac resynchronization therapy (CRT) is feasible using combined Micra and WiSE-CRT systems. This combination effectively reduces QRS duration and improves ejection fraction in heart failure patients.
Area of Science:
- Cardiology
- Medical Devices
- Electrophysiology
Background:
- Conventional cardiac resynchronization therapy (CRT) systems face limitations in certain patient populations.
- Leadless pacing systems like Micra offer advantages but may require upgrade for CRT.
- Heart failure patients with specific pacing needs or contraindications for traditional CRT benefit from innovative solutions.
Purpose of the Study:
- To assess the technical feasibility, safety, and efficacy of combining Micra and WiSE-CRT systems for totally leadless CRT.
- To evaluate acute QRS duration reduction and procedural success.
- To explore mid-term outcomes including ejection fraction and heart failure class.
Main Methods:
- Eight patients indicated for Micra upgrade or unsuitable for conventional CRT underwent implantation of the WiSE-CRT system synchronized with an existing Micra pacemaker.
- Technical success was defined by successful WiSE-CRT implantation and synchronized pacing.
- Primary endpoints included acute QRS duration reduction and freedom from major adverse events.
Main Results:
- All eight WiSE-CRT devices successfully detected Micra pacing and delivered synchronized left ventricular (LV) pacing.
- Significant acute reduction in QRS duration was observed (204.38 ms to 137.5 ms, P=0.012).
- At 6 months, a significant increase in LV ejection fraction (28.43% to 39.71%, P=0.018) was noted, without significant changes in LV reverse remodeling or NYHA class.
Conclusions:
- The combination of Micra and WiSE-CRT systems enables successful totally leadless CRT delivery.
- The WiSE-CRT system offers a unique solution for upgrading Micra patients to CRT capability without device replacement.
- Future advancements in leadless pacing may further expand the application of this combined approach.